I. Main models
SVGB Series Stainless Steel Vacuum Glove Box Parameter List
SVGB series vacuum glove box |
SVGB-1 |
SVGB-2 |
SVGB-3 |
Shape (L*W*H) |
1030*580*540 |
1320*750*690 |
1755*800*1024 |
Box (length L width W height H thickness T) |
600*500*500*5 |
800*650*650*5 |
1200*700*950*6 |
Net space inside the box (upper) |
460*340*490 |
660*460*640 |
1060*563*938 |
Net space inside the box (lower) |
460*490*490 |
660*640*640 |
1060*688*938 |
Window glass (L*H*T) |
548*243*19 |
748*373*19 |
1148*435*19 |
Bracket (L*W*H1) |
600*500*883 |
800*650*858 |
1200*700*808 |
Transition cabin |
φ200*280 |
φ273*350 |
φ350*400 |
Left side door opening (optional) |
330*260 |
360*480 |
450*600 |
Transition compartment drawer (optional) |
126*250 |
202*320 |
282*350 |
Tool compartment (optional) |
/ |
/ |
150*300 |
Glove hole diameter |
φ150 |
φ150 |
φ196 |
Glove material and size |
Rubber φ160*700 |
Rubber φ160*700 |
Rubber φ180*750 |
Center distance between two glove holes |
300 |
400 |
450 |
Glove hole height from the bottom of the box |
125 |
150 |
200 |
Box working pressure range |
-0.1—0 MPa |
-0.1—0 MPa |
-0.1–0 MPa |
Leakage rate |
-0.1MPa Holding pressure for 12 hours, temperature constant pressure increase <0.003MPa=3KPa (Level 2 per 1SO10648) |
Optional
Fully automatic humidity system |
Fully automatic control of the humidity/moisture inside the chamber, out of the humidity blindness. |
Fully automatic pressure system |
Constant pressure inside the chamber, especially for hazardous powder leakage. |
Ovens |
If vacuum baking is required, you can select the over-bin heating mode |
Purification system |
After vacuuming the inert gas, the water oxygen value is usually around 200PPM. Purification system, can achieve water oxygen value <1PPM |
Dedusting system |
Internal purification level, class 100/class 1,000, with optional dust removal system |
II. Product introduction
Vacuum glove box, also known as vacuum inert gas glove box, in the process of scientific experiments, some substances in the atmospheric state is very easy to oxidize and deliquescence, which makes the chemical reaction of the test substance and the sample of the front-end processing is very difficult, affecting the test process and test results. Vacuum glove box (also known as vacuum inert gas operation box) effectively solves these problems, the device can make the test samples can be safely put in and take out, and in the state of oxygen and water free operation, reaction and testing, to ensure the normal conduct of scientific experiments. This product is an ideal device for scientific experiments in colleges and universities, scientific research units and enterprise laboratories, and is widely used in biochemical, metallurgical, electronic, chemical, geological, mineral, pharmaceutical and other industries.
Stainless steel vacuum glove box is a high-purity inert gas filled into the box, and the cycle of filtering out the active substances in the laboratory equipment, also known as the glove box, inert gas protection box, dry box and so on.
III. Product structure principle
Stainless steel vacuum glove box is mainly composed of the main box, transition chamber and frame, such as users have other special requirements, can also be designed or modified according to the needs; box on the two to three glove operation interface, the operating box to meet the face of the toughened glass, a wide perspective without dead ends, the operator can clearly observe the operation process in the box, so that the process of the operation of the more intuitively displayed in front of the operator's eyes. Box, the transition chamber has two vacuum ball valves for gas in and out of the box, box on the reserved KF40 vacuum connector, the user can be used according to need, in the need for water or gas can be accessed from this situation. The box has built-in lighting and multi-hole socket.
IV. Product features
1. The whole box is made of stainless steel, and made of reinforced treatment, the surface plastic spraying, so solid corrosion-resistant and easy to clean;
2. The operation box is made of plexiglass, with wide view angle and no dead angle;
3. Inside the box with lighting and sockets;
4. The operating gloves are made of imported butyl gloves, which are reliable and durable;
5. The sealing device is novel in design, reliable sealing and easy to open;
6. The box is reserved for KF40 vacuum connector, which can be transformed and used according to the needs;
7. The two doors of the transition room solve the problem of repeated vacuuming of the main box.
V. Product use precautions
1. Before use, you must check whether the connections are loose and whether the seals are intact;
2. It is forbidden to open the hatch, glove port cover or disassemble and repair under vacuum;
3. The glove port should be vacuumed firstly or simultaneously with the box, and then stop vacuuming with the box at the same time or at the same time;
4. The connection of external gas source pipe fittings should be sealed securely, and the inlet valve should be closed when the needle of vacuum pressure gauge points to 0 when the box is inlet;
5. Close the valve and then close the vacuum pump after vacuuming;
6. Check whether the glove valve is open when pumping.