Lab 1600C High Vacuum High Temperature Microwave Sintering Furnace
Description
HAMiLab-HV is SYNOTHERM standardized most advanced high vacuum multi-function lab furnace. Besides of the entire ranges of features of Hamilab-V, It can achieve the microwave vacuum sintering systhesis, offer more advanced atmosphere control for air sensitive samples. Materials may be processed up to 1600 degree Celsius in a vacuum below 10-3Pa or in nitrogen or inert gases.
Features
- Adopt the industry-grade microwave source to ensure the furnace’s long time continuous steady running.
-Microwave output power adjustable freely, no gear limitation, to achieve accurate temperature control curve;
-Set with two-stage vacuum pump two gas passages (input output), it can offer accurate atmosphere control high vacuum for the materials sample.
-Adopt imported high-precision IR thermometer to measure directly the temperature of the sample.
-Be equipped with embedded micro-computer control system, offer manually operation, automatically operation, constant temperature operation three modes, shifted freely.
-All kinds of original crucibles insulation structures available for choice, no pollution to materials.
-It can deal with different kinds of materials, with good commonality;
-Set with the corrosion-resistant exhaust passage, exhaust rapidly the gas produced in the heating process.
-Touch screen display operation, equipped with windows operation system, able to be connected with PC; Capable of set store many temperature control curves. Store automatically running records, compatible with printer; real time display the temperature control curve, to achieve the dynamic monitor of the heating process;
- Safe reliable microwave shield design, multi-ply protection from microwave leakage.
Application Field
Besides of the entire application area of HAMiLab-V atmosphere furnace, it can also be applied to fhe following areas:
Iron-based cooper-based powder metallurgy Kentanium’s sintering
Neodymium iron boron etc magnetic materials’ sintering
Metal powder materials’ nitridizing carbonizing
Technical specification
Technical specification |
||
Model |
HAMiLab-HV3000 |
HAMiLab-HV6000 |
Microwave Output Power |
0.1 - 2.85kW variable |
0.3 – 5.7kW variable |
Working Voltage & Frequency |
AC 380 plus/minus 10V, 50Hz |
|
Microwave Frequency |
2.45 GHz plus/minus 25MHz |
|
Max Working Temperature |
1600 Celsius degree |
|
Temperature Measuring |
Raytek IR pyrometer, 450-2250℃ |
|
Temperature Control Precision |
0.3% ± 1℃ |
|
Static Vacuum Degree |
less than 6.67x10-3Pa |
|
Atmosphere System |
air, oxygen, nitrogen, argon, weakly reducing atmospheres, etc |
|
Chamber Dimensions (mm) |
Φ500 × 560 (Dia. × L) |
|
Atmosphere Control |
2 channels (optional) |
|
Effective Heating Zone (mm) |
Φ125x100 (Dia. × L) |
|
Cooling |
Water |
|
Hardness of Cooling Water |
less than 60mg/L |
|
Control System |
PLC (18 sections of temperature control), embedded system connected to laptop via USB, with data storage printing functions. |
|
Control Panel Display |
8” touch-pad LCD or laptop |
|
Protection Safety |
Anode over voltage, anode over current, anode power surpass, electromagnet watchdog, magnetron temperature, external interlock. |
|
Microwave Leakage |
< 1mW/cm2 (Safety standard < 5mW/cm2) |
|
Site area |
2m2 |